Model Number PMX110
Medical Device Problem Codes Decrease in Suction (1146); Device Emits Odor (1425)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the iliac vein using a penumbra system aspiration pump max 110 (pump max).During the procedure, the maximum vacuum pressure that the pump max generated was -22 mmhg.It was reported that this was not enough vacuum pressure to aspirate the thrombus; therefore, the pump max was disconnected.It was also reported that the pump max emitted a burning smell.The procedure was completed using a new pump max.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Additional information was added to: section e.Box 1: first name and postal code (zip).Results: the pump max was plugged in and powered on.The pump max was able to achieve a maximum vacuum pressure of -25in.Hg on the pump max gauge and the calibrated gauge.The pump max was powered on for 30 minutes and no burning smell was observed.Conclusions: evaluation of the returned pump max revealed a functional device.During the functional test, the pump max was plugged in and powered on.The pump max was able to achieve a vacuum pressure within specification.Therefore, the root cause of the reported complaint could not be determined.Penumbra pumps are functionally tested during incoming inspection by quality.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9801880
Report Number3005168196-2020-00304
Device Sequence Number13613756
Product Code DXE
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (Y/N)Y
Initial Reporter CountryCO
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMX110
Device Catalogue NumberPMX110
Device Lot NumberF13826-43
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/18/2019
Supplement Date Received by Manufacturer05/18/2020
Initial Report FDA Received Date03/06/2020
Supplement Report FDA Received Date06/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age27 YR
Date Report Sent to FDA01/01/2005
-
-