| Model Number |
MMT-7020A |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic the user reported inaccurate sensor readings that triggered a threshold suspend.The customer¿s blood glucose was 185 mg/dl and sensor glucose was 80 mg/dl at the time of the event, the difference is outside the acceptable range.Suspend on low limit in sensor settings was at 100 mg/dl.Sensor glucose and blood glucose difference was 95.Customer¿s other blood glucose level was 215 mg/dl and the other sensor glucose level was 120 mg/dl.No harm requiring medical intervention was reported.The sensor will not be returned for analysis.
|
| |
|