|
Catalog Number UNKNOWN |
Device Problems
Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Abdominal Pain (1685); Perforation of Vessels (2135); Visual Impairment (2138); Discomfort (2330); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
Manufacturer ref# (b)(4).Catalog# is unknown but referred to as cook gunther tulip filter.Occupation: non-healthcare professional.Pma/510(k) k172557.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook gunther tulip filter on an unknown date.It is alleged that the [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.".
|
|
Manufacturer Narrative
|
Device code(s): appropriate term/code not available (3191) was selected for the alleged device perforation.Appropriate term/code not available (3191) was selected for the alleged device tilt.Investigation is reopened due to additional information provided.The following allegations have been investigated: vena cava (vc) perforation, tilt the reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Catalog number and lot number are unknown; however, there is no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
(b)(6) 2020, per a report from computed tomography (ct); ¿findings: ivc: an ivc filter is present.There is tilt of the apex measuring 11 degrees to the right.The apex of the filter is touching the ivc wall.The apex of the filter is at the orifice of an accessory right renal vein which is 31 mm below the right main renal vein.A strut at the 1:00 position appears perforated approximately 3mm.A strut at the 3:30 position appears perforated approximately 12mm.A strut at the 6:00 position appears perforated proximally 4mm.There is no significant stenosis of the ivc.Here is no significant filter migration.There are no fractured or bent strut fragments.¿.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2009 due to blood clot prevention, pulmonary embolism (pe), and deep vein thrombosis (dvt).Patient is alleging tilt, shortness of breath, vision impairment or loss, back pain, abdominal pain, internal discomfort.
|
|
Manufacturer Narrative
|
Investigation: the following allegations have been investigated: shortness of breath, vision impairment/loss, back/abdominal pain, internal discomfort.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Unknown if the reported shortness of breath, vision impairment/loss, back/abdominal pain, and internal discomfort are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown; however, there is no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|
|
|