|
Visual inspection of the returned guidewire confirmed that the distal tip section was separated in two sections.The tip's corewire fractured at approximately 2.9 cm from the proximal end of tip.The combined length of the proximal (2.9cm) and distal (0.1cm) fragments confirmed that no part was left behind in the patient.The coil surrounding the corewire was stretched to almost its full length.Fault protection failed - the coil did not prevent separation of the corewire fragment.The characteristics of the fracture indicated that the guidewire broke as a result of pull force.The corewire displayed no sign of torsion.No evidence of material defect or of manufacturing issue was observed.Device history record (dhr) was reviewed and validated that the optowire iii lot was released per specifications, which include two 5n tensile strength tests (corewire and assembled tip) and final inspection on 100% of the devices prior to release.Review of records also confirmed that all outer diameter measurements of the tip section, performed on 100% of the devices, were within specifications (0.0140''/0.356mm guidewire).Inspection with electronic microscope of the distal segment of the guidewire validated that the tip outer diameter ([1] 0.343 mm/0.0135''; [2] 0.345mm/0.0136'') was conform to specifications.Device inspection, review of narrative and review of dhr did not identify any evidence that use of a damaged guidewire contributed to the incident.Vasoconstriction is hypothesized as a potential cause of the entrapment.Review of complaint data was performed - the event is the first documented case of tip fracture for the optowire iii.The investigation concludes that the tip of optowire iii was entrapped into a side branch of the lad.The corewire was subsequently pulled to break in the attempt to free the entrapped wire.Use error or device defect are not considered to have contributed to the event.The investigation did not identify any new risk.The risks associated with the event are disclosed in the optowire iii instructions for use (ifu): when taking the optowire out of the dispenser hoop, do not handle the optowire roughly or pull it out abruptly as this can damage the optowire.Inspect the optowire for bends, kinks or other damage prior to use.Do not use damaged optowire.Using a damaged optowire may cause vessel damage and/or inaccurate torque response and pressure reading.When shaping the distal end, do not shape with a sharp-edged instrument.Use the minimum force needed so that the coil is not damaged.Inspect the coil and optowire for damage after shaping and before using.Optowire should be manipulated only under fluoroscopy.Care should be taken when manipulating a guidewire inside a vessel during device placement and removal.Observe optowire movement in the vessels.Before an optowire is moved or torqued, the tip movement should be examined under fluoroscopy.Do not torque an optowire without observing corresponding movement of the tip; otherwise, vessel trauma may occur.Never advance an optowire against resistance without first determining the reason for the resistance under fluoroscopy.Excessive force against resistance may result in damage to the wire and/or to the vessel.If resistance occurs and the cause of resistance cannot be determined, do not move or torque the optowire.Stop the procedure, determine the cause of resistance under fluoroscopy and take appropriate action.No additional action is currently intended by opsens as a result of the incident.The potential reoccurrence of similar events will be monitored through opsens complaint handling process.The need for additional actions will be reassessed accordingly.
|