Patient sid: (b)(6).All available patient information were included, no additional patient information available.Correction/removal number = 3005094123-10/22/18-001-c.Product correction letters were issued on 12oct2018 to all architect total t3 (tt3) and free t3 (ft3) customers, using the i1000sr, i2000, or i2000sr instruments and who received one of the impacted lots.The letter instructed the customers to prevent this interaction by either separating the architect ft3 and/or architect tt3 assays by running these tests on different instruments from the following assays architect tsh, architect t-uptake, architect hiv ag/ab combo, architect cortisol, architect lh, architect progrp, architect rhtlv-i/ii, architect total psa, architect afp, and architect free psa, architect 25-oh vitamin d, or by performing daily maintenance on the instrument prior to performing batch testing for all ft3 and/or tt3 samples.The mechanism of carryover involves carryover of poly-l-lysine (pll) from other assays into the conjugate dispense step in the ft3 and tt3 assays through adherence of the pipetting probe.
|
|
The customer observed bio-rad quality controls are out of range low and results were false depressed on 2 patients for freet3(ft3) while processing with the reagent lot 06426ui00 on the architect i2000sr.The following patient data was provided.Sid (b)(6); initial result 1.78 pg/ml, repeat on the reference instrument 2.53 pg/ml, sid (b)(6); initial result 1.85 pg/ml, repeat on the reference instrument 2.56 pg/ml, the customer uses normal range 1.88-3.18.The customer stated results are within range with the reagent lot 0402iu00.There were no patient results provided.There was no reported impact to patient management.
|