This event was initially reported under mfr # 2916596-2017-02059 and mfr # 2916596-2017-02057 manufacturer's investigation conclusion: the evaluation of heartmate ii lvas, serial number (b)(4), did not reveal any device-related issues.The patient was found to have a match for a donor heart and underwent a heart transplant on (b)(6) 2017.The pump was returned assembled with the driveline (dl) cut approximately 6¿ from the pump housing and the remaining distal portion of the dl was not returned.The inflow conduit (inlet tube, flex section, and inlet elbow), outflow graft, outflow graft bend relief, and bend relief collar were not returned.Outflow elbow was returned attached to the pump¿s outlet port.Evaluation of the outflow elbow revealed no evidence of depositions or thrombus formations.Visual examination of the pump¿s blood-contacting surfaces upon disassembly of (b)(4) also revealed no evidence of depositions or developed thrombus formations.The disassembled pump¿s bearings, rotor, and blood-contacting surfaces were examined under a microscope and no anomalies were observed.Electrical continuity testing of the driveline did not reveal any discontinuities or shorts.The pump was cleaned, reassembled and functionally tested under normal operating conditions using a mock circulatory loop.The data retrieved from that testing revealed normal pump power consumption and pressure values, comparable to what was recorded during the manufacturing process and the device functioned as intended.No further information was provided.The manufacturer is closing the file on this event.
|