|
During preparation for the procedure, when the laser fiber was removed from the sterile packaging, it was noted the fiber was kinked/ fractured.The device was set aside and a new of the same device was used to successfully complete the procedure.There was no harm or injury to the patient due to the event as the device did not come into contact with the patient.It was reported the disposable device is available for return to the manufacturer for a device evaluation.
|
|
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.As a device evaluation could not be performed, a root cause for the event cannot be determined.The customer's reported complaint of a fractured fiber could not be confirmed as the sample was not returned for evaluation.At the vendor facility, manufacturing and inspections procedures include a 100% inspection of the entire fiber, including the quality of each strip.Each unit is also repeatedly bent and coiled during the processing and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to packaging and shipment.During the hene laser test the fiber tip is examined closely for damage.During the packaging step, the fibers are inspected for damage.Without receiving product to evaluate, the root cause cannot be determined, although it is unlikely that the fiber was fractured prior to packaging.The most likely root cause to this event is due to handling after the device left the angiodynamics facility.A review of the angiodynamics' device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The directions for use, which is supplied to the end user with this catalog number, contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference complaint (b)(4).
|
|
Returned for evaluation was a 55 ops fiber.A visual examination of the device noted a fracture 93cm from the tip.The customer's reported complaint description of the fiber fracturing is confirmed.A review of the returned sample shows the shaft of the fiber shows flaws within the glass core which can impact the resultant fiber strength.A definitive root cause cannot be determined, a similar event occurred (reference (b)(4)): "the root cause was most likely due to an inherent flaw in the fiber glass core that failed while the fiber was packaged in a coiled configuration".As stated in the ifu "avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a diameter of 16cm." a lot history records search for the nevertouch kit revealed no quality related issues or manufacturing deficiencies at the time of manufacture.During the manufacturing process, the disposable fiber device receives a 100% inspection and an aql inspection in which the quality of the fiber strip is inspected.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint reference (b)(4).
|