| Model Number |
110-090 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
12/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture baker grade iii is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Event or Problem Description
|
|
Health professional reported right side capsular contracture, baker grade iii.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: visual analysis of the returned device identified: crease fold, deformation, yellow particles in the surface of the shell and weight to the spec.Cloudy and voids was observed after autoclave disinfection process.Based on the device analysis the final assessment is: - no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Health professional reported right side capsular contracture, baker grade iii.The device has been explanted.
|
| |
|