C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3 GO¿ INTERMITTENT URINARY CATHETER; MAGIC 3 MALE COUDE Back to Search Results
Model Number 50818G
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Codes Injury (2348); Patient Problem/Medical Problem (2688)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the patient stated the catheters scraped him upon insertion.No medical intervention reported.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential failure mode could be ¿loose mandrel guides/excess vibration on the pallet¿ with a potential root cause of ¿mandrels touching each other within the pallet¿.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿review the self-catheterization procedure with a healthcare professional." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the patient stated the catheters scraped him upon insertion.No medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3 GO¿ INTERMITTENT URINARY CATHETER
Common Device Name
MAGIC 3 MALE COUDE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9805889
Report Number1018233-2020-01652
Device Sequence Number2909672
Product Code EZD
UDI-Device Identifier00801741137044
UDI-Public(01)00801741137044
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K172247
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number50818G
Device Catalogue Number50818G
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/17/2020
Supplement Date Received by Manufacturer04/08/2020
Initial Report FDA Received Date03/09/2020
Supplement Report FDA Received Date04/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-