Product complaint # (b)(4).Investigation summary: update 20-feb-2020: the investigation was re-opened upon the receipt of the product.Examination of the returned device found the wrench to be difficult to operate.The noted damage is consistent with excessive torque being used and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: update 20-feb-2020: the investigation was re-opened upon the receipt of the product.Examination of the returned device found the wrench to be difficult to operate.The noted damage is consistent with excessive torque being used and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|