• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 6IN AML SM STATURE 12.0MM; AML IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 6IN AML SM STATURE 12.0MM; AML IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 1554-01-121
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); Weakness (2145); Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 08/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Update oct.5, 2017: litigation record received.Litigation alleges pain, weakness and difficulty with simple living activities.Update feb.21, 2020: after review of medical records, the patient was revised due to wear of articular weight-bearing surface, pain and loosening of the femoral stem.Radiographs showed no signs of loosening and the stem appeared to be spot-welded in several spots.The patient's cobalt chromium levels were elevated.Operative notes indicate that there was a lot of scarring, the stem was rotationally loose.Doi: (b)(6) 2010; dor: (b)(6) 2017; right hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6IN AML SM STATURE 12.0MM
Common Device Name
AML IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9808392
Report Number1818910-2020-07279
Device Sequence Number654045
Product Code LPH
UDI-Device Identifier10603295058236
UDI-Public10603295058236
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K061833
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2017
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/18/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1554-01-121
Device Catalogue Number155401121
Device Lot NumberEA6CC1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/21/2020
Supplement Date Received by Manufacturer04/07/2020
Initial Report FDA Received Date03/09/2020
Supplement Report FDA Received Date04/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
-
-