• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008K2 MACHINE; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008K2 MACHINE; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number UNKNOWN-2008K2 MACHINE
Device Problem Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the device was not returned to the manufacturer for physical evaluation and the serial number could not be obtained.As a serial number could not be determined, device history and manufacturing records could not be reviewed.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.Should additional relevant information become available, a supplemental report will be submitted. .
 
Event Description
A biomedical technician (biomed) reported noticing that a 2008k2 machine had internal wiring insulation that was breaking off and displayed signs of melting.The biomed stated that this issue could lead to conductivity and temperature issues, over time.To resolve the reported issue, the biomed replaced the conductivity sensor, and the two temperature sensors (temperature sensor 2 and temperature sensor 3).The wiring from all three components was exposed.There was no patient involvement associated with the reported issue.No sample was available to be returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2 MACHINE
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key9809216
MDR Text Key193640039
Report Number2937457-2020-00449
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN-2008K2 MACHINE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received02/18/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-