| Model Number |
401575 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/03/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission. .
|
| |
|
Event or Problem Description
|
|
During the procedure, the catheter became entrapped in a posterior branch of the eyelet vein.Due to difficult device removal, the catheter was left in the patient for the remainder of the procedure.At the end of the case, the device was still stuck so the deflection mechanism was cut off to allow a sheath to advance over the catheter shaft.A dose of verapamil was administered and the catheter was removed successfully.Fluoroscopy was used with contrast to observe the vein but the vein was functioning properly and no damage was noted.The physician did not believe the event was due to a performance issue with the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d10, g4, g7, h2, h3, h6, h10.The reported entrapment issue was unable to be confirmed due to the condition of the returned device.The shaft had been cut off distal to the handle and the handle was note returned.Functional testing was not possible due to the aforementioned damage.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported entrapment may have been procedure related.
|
| |
|