TITAN SPINE ENDOSKELETON TCS; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL
|
Back to Search Results |
|
| Model Number |
5302-3514 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient identifier and patient weight were not available.The subject device has not been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a spinal device being used for a 3 level anterior cervical discectomy and fusion (acdf) case.The acdf procedure was being performed to treat spinal stenosis, and the levels implanted were c4/5, c5/6, and c6/7.It was reported that during the procedure the locking screw's locking cap sheared off the screw.The surgeon tried inserting the locking screw into the c4/5 body.After using a curved free hand awl into the device, the surgeon drove the locking screw through the spacer into the bone.Upon reaching the lip of the spacer, the locking cap sheared off the screw and into the incision site.Everything was recovered, and the cap and screw shaft were removed from the patient, confirmed by fluoroscopic imaging.No further problems arose thereafter.The curved variable awl was used to make a new trajectory, and a new screw was inserted into the new trajectory.There were no patient symptoms or complications, and there was no additional treatment or surgery performed as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|