ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1120-040
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was received.The investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat an av fistula.The 12x40mm armada 35 balloon dilatation catheter (bdc) was advanced to the lesion with no issues.During the first inflation, the balloon ruptured at the rated burst pressure (rbp).During retraction, resistance was met with the anatomy and the balloon separated.A snare device was used to retrieve the separated portion from the anatomy.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
Visual analysis was performed on the returned device.The reported balloon rupture and separation was confirmed.The difficulty removing was not tested due to the condition of the device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product issue.The investigation determined that the reported difficulties were likely due to case related circumstances.It is likely that the balloon rupture initiated longitudinal and was the result of interaction with lesion calcified or associated devices.In addition, the resistance noted during removal and radial separation of the balloon was likely the result of the ruptured balloon material catching on the introducer sheath during removal.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9811986
Report Number2024168-2020-02203
Device Sequence Number11493377
Product Code LIT
UDI-Device Identifier08717648154980
UDI-Public08717648154980
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model NumberB1120-040
Device Catalogue NumberB1120-040
Device Lot Number90628G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2020
Initial Date Received by Manufacturer 02/18/2020
Supplement Date Received by Manufacturer03/10/2020
Initial Report FDA Received Date03/10/2020
Supplement Report FDA Received Date03/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-