MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN RELIAMAX; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN RELIAMAX; STAPLE, IMPLANTABLE Back to Search Results
Model Number GIA8038RL
Medical Device Problem Codes Misfire (2532); Failure to Form Staple (2579); Detachment of Device or Device Component (2907); Fail-Safe Problem (2936)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/30/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during hemicolectomy with total abdominal hysterectomy and salpingooophorectomy, after opening the sterile packaging, yellow shipping wedge was already removed without even removing it manually.When applied on the anastomosis, the firing knob broke for the white flag interlock already engaged.It was also reported that device partially fired and had poor staple formation which staples got dismantled.Misfired staples fell into the patient cavity.Surgeon oversewed to fix the staple line.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIAMAX
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key9811995
Report Number1219930-2020-01098
Device Sequence Number14154986
Product Code GDW
UDI-Device Identifier20884521168630
UDI-Public20884521168630
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K122781
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model NumberGIA8038RL
Device Catalogue NumberGIA8038RL
Device Lot NumberN8M0794Y
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2020
Initial Report FDA Received Date03/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-