| Model Number |
EPM8745 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cabg heart procedure on (b)(6) 2020 and suture was used.During the procedure, while suturing the distal portion of the anastomosis, the suture broke at the mid-point.Another like device was used to complete the procedure with no adverse patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent to the fda: 04/07/2020.A manufacturing record evaluation was performed for the finished device lot: pgr878, and no non-conformances were identified.Additional h3 investigation summary: it was received for analysis five unopened samples of product code: epm8745, lot: pgr878.During the visual inspection of five unopened samples, no defects were found on the packages.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the samples received, no performance - breakage suture was found and the tested samples met the finished goods requirements.
|
| |
|
Search Alerts/Recalls
|