• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT300
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, information not provided.If explanted, give date: not applicable due to lens has not been explanted.Device evaluation: product testing could not be performed as the lens remains implanted, the product was not returned.The reported complaint could not be verified.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no other complaints were received for this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.Note: the device was manufactured at the kulim site which has been closed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zct300 22.5 diopter intraocular lens (iol) likely rotated greater than 10 degrees and has causing the patient to experience blurry vision.The doctor plans on explanting the iol.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: section d10: device available for evaluation, yes.Section d10: returned to manufacturer on, 03/16/2019.Section h3: device returned to manufacturer: yes.With the return of the device explant of the device the planned explant has taken place, therefore section b2 is now updated to required intervention, and patient code 3191-explant is being entered.Section b2: required intervention to prevent permanent impairment/damage.Section h6: patient code 3191 - explant.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the return condition of the lens no product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency was identified.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9815674
Report Number2020664-2020-00020
Device Sequence Number16911379
Product Code HQL
UDI-Device Identifier05050474553378
UDI-Public(01)05050474553378(17)210703
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/03/2017
Device Model NumberZCT300
Device Catalogue NumberZCT300U225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/11/2020
Supplement Date Received by Manufacturer03/19/2020
10/25/2020
Initial Report FDA Received Date03/11/2020
Supplement Report FDA Received Date03/20/2020
11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Date Report Sent to FDA01/01/2005
-
-