| Model Number |
MMT-551LNAL |
| Medical Device Problem Codes |
Power Problem (3010); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump was turned off with a battery change.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device monitored for several days.Device received with all operating currents within specification and passed a21 error test and displacement test.No unexpected battery out limit alarm anomalies noted.Device received with cracked reservoir tube lip and cracked battery tube threads.The test infusion set and reservoir does lock into place.(b)(4).
|
| |
|