| Model Number |
89-9429 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Sparking (2595)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: an internal complaint ((b)(4)) was received indicating a hand cautery pencil included within a convenience kit (part 89-9429, lot 5110378) was sparking during use.A sample was not available for evaluation.The work order for the reported finished good lot was reviewed for any discrepancies that may have contributed to the reported event.No discrepancies were identified.The bill of materials for the convenience kit was reviewed and the affected component was identified as raw material (b)(4).This is a cautery pencil supplied to deroyal by conmed.A supplier corrective action request (scar) was issued to conmed.As of the date of this report, a response has not been received.The investigation is ongoing.When new and critical information becomes available, this report will be updated.
|
| |
|
Event or Problem Description
|
|
The hand cautery in a convenience kit sparked during use.
|
| |
|
Event or Problem Description
|
|
The hand cautery in a convenience kit sparked during use.
|
| |
|
Additional Manufacturer Narrative
|
|
Root cause: the cautery pencil contained in the convenience kit is supplied to deroyal by conmed.Therefore, a supplier corrective action report (scar) was issued to conmed.In its response, conmed stated it was unable to identify a root cause for the reported issue because a sample was not available for evaluation.Corrective action: a new electrical test was implemented (b)(6) 2019 as part of an improvement.The reported lot number for the cautery pencil was prior to the implementation of this improved electrical test.Investigation summary: an internal complaint (b)(4) was received indicating a hand cautery pencil included within a convenience kit (part 89-9429, lot 5110378) was sparking during use.A sample was not available for evaluation.The work order for the reported finished good lot was reviewed for any discrepancies that may have contributed to the reported event.No discrepancies were identified.The bill of materials for the convenience kit was reviewed and the affected component was identified as raw material 709124.This is a cautery pencil supplied to deroyal by conmed.A supplier corrective action request (scar) was issued to conmed.A response was received and accepted april 13, 2020 by deroyal personnel.The investigation is complete.If new and critical information becomes available, this report will be updated.
|
| |
|
Search Alerts/Recalls
|