Model Number 5MAXJETDKIT
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional 510(k)# that also applies to this complaint: (b)(4).The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2020-00315.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure to treat a supraclinoid carotid occlusion using a penumbra system 3max reperfusion catheter (3maxc), a penumbra system jetd reperfusion catheter (jetd), and a neuron max 6f 088 long sheath (neuron max).It was noted that the patient anatomy was tortuous.During the procedure, the physician advanced the neuron max into the patient, then used the 3maxc to bring up the jetd to the treatment site through the neuron max.While attempting to remove the 3maxc, the physician experienced resistance.The physician then pulled harder and noticed that the 3maxc was stuck in the jetd.The physician then pulled even harder and was able to remove the 3maxc.However, upon removal, it was noticed that the 3maxc was broken and only the proximal end of the 3maxc was removed.The physician then removed the jetd and noticed that the broken distal end of the 3maxc was inside the jetd.The procedure was completed using a microcatheter, a stent retriever and the same neuron max.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM JETD REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9818474
Report Number3005168196-2020-00316
Device Sequence Number11314023
Product Code NRY
UDI-Device Identifier00814548019802
UDI-Public00814548019802
Combination Product (Y/N)Y
Initial Reporter CountrySP
PMA/510(K) Number
K090752
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/10/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2022
Device Model Number5MAXJETDKIT
Device Catalogue Number5MAXJETDKIT
Device Lot NumberC17810
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2020
Initial Report FDA Received Date03/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age41 YR