After review of medical records patient was revised to addressed painful left total hip arthroplasty with metal on metal articulation and increased serum metal ions.Operatives notes indicated small amount of black corrosion on the trunnion, noted some small scoring on the anterior neck of femoral component and this corresponded to some impingement on the anterior cup.Doi: (b)(6) 2008, dor: (b)(6) 2015, left hip.
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|