• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 9 HI OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 9 HI OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-14-090
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Granuloma (1876); Pain (1994); Visual Impairment (2138); Discomfort (2330); Osteolysis (2377); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Date of Event 12/04/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pcf and medical records received.After review of medical records, patient undergone physical examination last (b)(6) 2017.Recent labs reveal cobalt increased from 20.2 to 25.Chromium is stable at 11.It was 10 before.The patient has significant functional deficits, recent visual problem, trochanter tenderness, pain in rotation of the leg, leg length discrepancy, and severe right hip stiffness.The patient ambulates with an antalgic trendelenburg gait and moves about at a moderate pace.Medical records also stated an osteolysis in the right acetabulum with metal on metal implant.Patient was revised to address right failed total hip arthroplasty, metal on metal reaction, osteolysis, granuloma, elevated ion counts, and mild taper corrosion.Operative notes stated no evidence of infection.Surgeon performed extensive debridement on granuloma tissue.Doi: (b)(6) 2009.Dor: (b)(6) 2017; right hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 9 HI OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9818750
Report Number1818910-2020-07477
Device Sequence Number11530307
Product Code KWA
UDI-Device Identifier10603295001119
UDI-Public10603295001119
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K073570
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012-14-090
Device Catalogue Number101214090
Device Lot NumberCN7DE1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/23/2020
Supplement Date Received by Manufacturer04/15/2020
Initial Report FDA Received Date03/11/2020
Supplement Report FDA Received Date04/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient Weight76
-
-