• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5532-G-611
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348)
Date of Event 02/12/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding infection involving a tibial insert was reported.The event was not confirmed.Method & results: product evaluation and results: not performed as no product was returned for evaluation.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.There have been no other similar events for the sterile lot referenced.Conclusion: infection is a known possible adverse outcome of surgery, as noted in the instructions for use.All stryker products sold as sterile are validated to a minimum sterility assurance level (sal) of 10^-6 in accordance to applicable iso standards.When the sterile barrier of any device is opened, the sterility of that device becomes a function of handling and surgical technique and is beyond stryker¿s control.Infection due to bacterial contamination, as supplied from the manufacturer, is an extremely rare event.If further information becomes available or the product is returned, this investigation will be re-opened.The following devices were also listed in this report: triathlon ps x3 tibial insert; cat# 5551-g-381; lot# rnmx.Triathlon ps fem comp #7r-cem; cat# 5515f702; lot# d6b4n.Tri ts baseplate size 6; cat# 5521-b-600; lot# dgu7ba.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Device not returned.
 
Event or Problem Description
It was reported that a stage 1 revision was performed on the patient's right knee due to infection.The patient's knee construct was removed and a spacer was placed.Rep reported that no further information will be available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9819529
Report Number0002249697-2020-00503
Device Sequence Number16911120
Product Code MBH
UDI-Device Identifier07613327050851
UDI-Public07613327050851
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number5532-G-611
Device Catalogue Number5532-G-611
Device Lot Number9M6PTE
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/12/2020
Initial Report FDA Received Date03/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age57 YR
Patient Weight157
-
-