Model Number 15
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Physio-control evaluated the customer's device and was able to duplicate the reported issue.Physio-control determined that the cause of the reported issue was due to the system pcb assembly.Due to a part obsolescence, the device cannot be repaired at this time.The customer has received a loaner unit until a permanent solution is identified.
 
Event or Problem Description
The customer contacted physio-control to report that their device would not complete the boot up cycle.In this state the device would not be able to deliver defibrillation therapy if needed.There was no patient use reported with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATIC EXTERNAL DEFIBRILLATOR (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key9820937
Report Number0003015876-2020-00352
Device Sequence Number19592286
Product Code MKJ
UDI-Device Identifier00883873979020
UDI-Public00883873979020
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-000093
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/19/2020
Initial Report FDA Received Date03/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-