The device was returned to stryker sustainability solutions for evaluation.Visual inspection revealed a kink 9.8 cm from the distal tip of the device which is distal to the transition point.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause is a kink as a result of mishandling, including shipping and storage conditions, subsequent to distribution from stryker.The instructions for use (ifu) state: do not introduce the tip folded into the guiding sheath.Do not exert excessive pressure during placement of catheter if unknown resistance is encountered.Do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.Avoid excessive torque, stretching, kinking and/or bending of catheter, as this may interfere with distal tip shaping or cause damage to internal electrode wires.Do not alter this device.The reported event will continue to be monitored through post-market surveillance.
|