STRYKER SUSTAINABILITY SOLUTIONS PHOENIX; CATHETER, RECORDING, ELECTRODE, REPROCESSED Back to Search Results
Model Number 402004
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Visual inspection revealed a kink 9.8 cm from the distal tip of the device which is distal to the transition point.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause is a kink as a result of mishandling, including shipping and storage conditions, subsequent to distribution from stryker.The instructions for use (ifu) state: do not introduce the tip folded into the guiding sheath.Do not exert excessive pressure during placement of catheter if unknown resistance is encountered.Do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.Avoid excessive torque, stretching, kinking and/or bending of catheter, as this may interfere with distal tip shaping or cause damage to internal electrode wires.Do not alter this device.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the catheter was bent.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS PHOENIX
10232 s. 51st st.
phoenix AZ 85044
MDR Report Key9821180
Report Number0002090040-2020-00010
Device Sequence Number1767367
Product Code NLH
UDI-Device Identifier00885825007744
UDI-Public00885825007744
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K012708
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/16/2021
Device Model Number402004
Device Catalogue Number402004RR
Device Lot Number4011607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2020
Initial Report FDA Received Date03/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-