According to the reporter, they had a capsule, which failed to attach.There was no harm to the patient and user.No intervention was required.There was nothing unusual about the patient or the procedure and an endoscopy had been performed prior to the procedure and showed the esophagus to be normal.Lubrication was used to facilitate placement of the capsule and the delivery system and capsule will be returned for investigation.
|
|
Additional information: b5, d4 (udi), d10, g4, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was returned for evaluation.It was reported that the capsule failed to attach to the patient's esophagus.The reported issue could not be confirmed.Visual inspection found no potentially contributing factors, and the sample met all related specifications.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that products meet all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The most likely root cause could not be identified because no problem was detected with the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|