| Model Number |
MMT-1760KPK |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a missing retainer ring.The customer stated that the reservoir does not lock into place, resulting in the reservoir falling out of the pump or the pump giving her the incorrect amount of insulin.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Tested with a test p-cap and the test p-cap did not lock in place properly due to missing retainer.Unable to perform displacement test due to missing retainer.Device received with partially broken off reservoir tube lip, missing reservoir tube o-ring, cracked select button on keypad overlay, pillowing keypad overlay, scratched/peeling keypad overlay texture, scratched display window cover, faded/stained/peeling serial number label, peeling/fading end cap address label, cracked battery tube area, cracked battery tube threads and scratched case.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0956-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|