• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MINI TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MINI TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012268-06U
Medical Device Problem Codes Material Rupture (1546); Material Separation (1562); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.(b)(4).
 
Event or Problem Description
It was reported the procedure was to treat a lesion in the left main (lm) artery.The 1.20x6mm rx mini trek balloon dilatation catheter (bdc) was advanced to the lesion and the balloon inflated three times to 20, 15, and 15 atmospheres (atm) respectively when the balloon ruptured.The device was removed however it was noted the shaft had separated, the distal 10cm of the device remained in the anatomy.Attempts to retrieve the separated portion were unsuccessful and the device remains in the anatomy.The patient was transferred to the intensive care unit (icu) to be further monitored.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned.Visual inspection was performed on the returned device.The reported separation was confirmed.The reported balloon rupture could not be determined since the balloon was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other similar incidents/complaints from this lot.It should be noted that the coronary dilatation catheters (cdc), mini trek instruction for use (ifu) states: balloon pressure should not exceed the rated burst pressure (rbp).The rbp for the mini trek device is 14 atm; therefore, rbp was exceeded.The investigation determined the reported balloon rupture appears to be related to user error as it was reported by the account that 1.20 x 6mm rx mini trek bdc was advanced to the lesion and the balloon inflated three times to 20, 15, and 15 atmospheres (atm) respectively when the balloon ruptured.The reported balloon separation, foreign body in patient, additional treatment and hospitalization appear to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9823562
Report Number2024168-2020-02311
Device Sequence Number1299574
Product Code LOX
UDI-Device Identifier08717648176531
UDI-Public08717648176531
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model Number1012268-06U
Device Catalogue Number1012268-06U
Device Lot Number90731G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2020
Initial Date Received by Manufacturer 02/27/2020
Supplement Date Received by Manufacturer05/11/2020
Initial Report FDA Received Date03/12/2020
Supplement Report FDA Received Date05/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-