Model Number 1012455-20
Medical Device Problem Codes Off-Label Use (1494); Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/18/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was received.Investigation has not yet been completed.A follow up report will be submitted with all additional relevant information.(b)(4).
 
Event or Problem Description
It was reported that this was a procedure to treat a lesion with heavy calcification and moderate tortuosity in the superior mesenteric artery (sma).The lesion was pre-dilated, and a 4.0 x 38 mm xience sierra stent was implanted successfully.Intravascular ultrasound (ivus) was performed and the stent was well apposed.A 5.0 x 20 mm nc trek balloon catheter was advanced for post dilation.However, at the third inflation at 16 atmospheres the balloon ruptured.During removal of the balloon catheter, the distal shaft separated in two pieces inside the guiding catheter as there was resistance between the devices.A snare device was used to securely with withdrawal the separated distal shaft.The patient remained stable.
 
Additional Manufacturer Narrative
Visual and functional inspections were performed on the returned device.The reported separation was confirmed.The reported balloon rupture was not confirmed; however, the noted tear in the outer member distal to the outer member to hypotube seal is likely what the account perceived as the reported balloon rupture.The reported difficulty removing the device could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty removing the device and balloon rupture; however, the reported separation and additional treatment appear to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK RX
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9824064
Report Number2024168-2020-02319
Device Sequence Number1299577
Product Code LOX
UDI-Device Identifier08717648152160
UDI-Public08717648152160
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number1012455-20
Device Catalogue Number1012455-20
Device Lot Number00109G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2020
Initial Date Received by Manufacturer 02/18/2020
Supplement Date Received by Manufacturer03/20/2020
Initial Report FDA Received Date03/12/2020
Supplement Report FDA Received Date04/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4X38MM XIENCE SIERRA STENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient Weight46
-
-