• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO PROTACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO PROTACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 174006
Device Problem Misfire (2532)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/03/2020
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection of the returned product noted that the timing was disrupted and a tack was protruding from the tip of the device.A functional evaluation found that the trigger was actuated and the remaining tacks deployed but did not seat properly due to the disrupted timing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition is caused by an instrument that has been exposed to excessive force while applying helixes to a surface.If a helix is fired over improper surfaces it can provoke the exertion of excessive force to the handle causing the unit to disrupt the timing and to a possible jam.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during mesh placement in a laparoscopic hernia repair procedure, the device misfired.Another device was used to complete the case.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTACK
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave.
north haven, CT 06473
2034925563
MDR Report Key9824128
MDR Text Key183125010
Report Number2647580-2020-00919
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10884521070400
UDI-Public10884521070400
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Model Number174006
Device Catalogue Number174006
Device Lot NumberP9A1159PY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-