Model Number 1516-40-810
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Edema (1820); No Code Available (3191)
Date of Event 10/07/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2017: patient underwent a right knee replacement due to djd.Attune implants were placed with depuy cement.On (b)(6) 2019: patient underwent a right knee revision due to partial debonding of the tibial tray.The femoral component was also found to have early loosening (unknown interface).There was slight internal rotation of the femoral component which cause slight lateral tracking of the patella.A large effusion (150mls) was found as well as global instability.Doi: (b)(6) 2017, dor: (b)(6) 2019, (right knee).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FB INSRT SZ 8 10MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9824195
Report Number1818910-2020-07617
Device Sequence Number16911241
Product Code JWH
UDI-Device Identifier10603295050384
UDI-Public10603295050384
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K111433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number1516-40-810
Device Catalogue Number151640810
Device Lot NumberH76569
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/21/2020
Supplement Date Received by Manufacturer04/07/2020
Initial Report FDA Received Date03/12/2020
Supplement Report FDA Received Date04/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 8 CEM; ATTUNE MEDIAL ANAT PAT 38MM; ATTUNE PS FB INSRT SZ 8 10MM; ATTUNE PS FEM RT SZ 8 CEM; SMARTSET HV BONE CEMENT 40G; SMARTSET HV BONE CEMENT 40G; ATTUNE FB TIB BASE SZ 8 CEM; ATTUNE MEDIAL ANAT PAT 38MM; ATTUNE PS FB INSRT SZ 8 10MM; ATTUNE PS FEM RT SZ 8 CEM; SMARTSET HV BONE CEMENT 40G; SMARTSET HV BONE CEMENT 40G
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-