SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 36MM +0; PROSTHESIS, HIP, SEMI-CONSTRAINED, MTAL/POLYMER, POROUS UNCEMENTD Back to Search Results
Model Number 71303600
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/21/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a thr procedure, the patient has undergone recurrent dislocations due to soft tissue laxity.For this reason, revision surgery was performed and the 36mm +0 cocr head was replaced with a 22mm +0 oxinium head ref 71302200 lot 18dm01388.All explanted implants have been discarded.
 
Additional Manufacturer Narrative
The devices, used in treatment, were not returned for evaluation, and the reported event could not be confirmed.A review of the device history records for the reported batches did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Although there were no prior complaints for three of the four reported devices.A clinical analysis indicated that it was reported that after a thr procedure (b)(6) 2018, the patient has undergone recurrent dislocations due to soft tissue laxity.For this reason, revision surgery was performed 17-feb-2020 and the stem was removed to gain access to the acetabulum (stem was well fixed, well-positioned) and replaced with a same-sized stem.All explanted implants have been discarded.No medical imaging or clinical documentation has been provided to confirm the recurrent dislocations.Without supporting medical documentation, a thorough medical assessment cannot be performed.In the event additional medical/clinical records are received, the clinical task will be re-opened and a thorough assessment will be rendered at that time.The risk management file and instructions of use for the product were reviewed.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction or loss of sterility.Based on this investigation, the need for corrective action is not indicated.If new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.Be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HEAD 36MM +0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, MTAL/POLYMER, POROUS UNCEMENTD
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9824335
Report Number1020279-2020-00843
Device Sequence Number3068030
Product Code LPH
UDI-Device Identifier03596010479983
UDI-Public03596010479983
Combination Product (Y/N)N
PMA/510(K) Number
K022902
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71303600
Device Catalogue Number71303600
Device Lot Number18EM18386
Initial Date Received by Manufacturer 02/21/2020
Supplement Date Received by Manufacturer11/11/2020
Initial Report FDA Received Date03/12/2020
Supplement Report FDA Received Date11/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-