• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS INC. PRELUDE IDEAL¿ HYDROPHILIC INTRODUCER SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS INC. PRELUDE IDEAL¿ HYDROPHILIC INTRODUCER SHEATH Back to Search Results
Model Number 10884450317485
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device has been returned for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that during vascular access the introducer hub detached.The sheath was exchanged for the procedure.No patient injury to report.
 
Additional Manufacturer Narrative
The suspect device has been returned for evaluation.The complaint is confirmed.The root cause is attributed to the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.Corrective actions are in process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRELUDE IDEAL¿ HYDROPHILIC INTRODUCER SHEATH
Common Device Name
INTRODUCER SHEATH
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan,, ut
MDR Report Key9826398
Report Number1721504-2020-00010
Device Sequence Number16468723
Product Code DYB
UDI-Device Identifier10884450317485
UDI-Public10884450317485
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K173750
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model Number10884450317485
Device Catalogue NumberPID6F11021SSC
Device Lot NumberH1347658
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2020
Initial Date Received by Manufacturer 02/14/2020
Supplement Date Received by Manufacturer03/06/2020
Initial Report FDA Received Date03/12/2020
Supplement Report FDA Received Date04/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-