| Model Number |
00884450038109 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/13/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a vascular procedure, the catheter tip detached within a patient.The physician had acquired arterial vascular access and during catheter manipulations the tip detached within the patient's common femoral artery.The physician used a vascular snare to successfully retrieve the catheter tip from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.The root cause was attributed to significant force being applied to the device during use.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed and no exception documents were found.
|
| |
|