• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA PLATELET SAMPLER, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA PLATELET SAMPLER, PLASMA, RBC SET Back to Search Results
Model Number 82440
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: w141619596804 the platelet collection set is not available for return because it was discarded by the customer.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h6 and h10.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.The run data file (rdf) was analyzed for this event.Signals from the rdf indicated there was potential for contamination just prior to a scheduled purge.Root cause: the analysis of the run data file did not find a conclusive cause for the higher-than- expected wbc content in the platelet product reported for this collection.Signals from the rdf indicate there was potential for contamination just prior to a scheduled purge.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA PLATELET SAMPLER, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key9827497
Report Number1722028-2020-00103
Device Sequence Number14106325
Product Code GKT
UDI-Device Identifier05020583824409
UDI-Public05020583824409
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2021
Device Model Number82440
Device Catalogue Number82440
Device Lot Number1908144251
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2020
Supplement Date Received by Manufacturer07/22/2020
Initial Report FDA Received Date03/12/2020
Supplement Report FDA Received Date07/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-