• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. MULTIFIX S-ULT 6.5MM KNOTLESS ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. MULTIFIX S-ULT 6.5MM KNOTLESS ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72290002
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that during shoulder arthroscopy procedure the anchor was without cut "blunt" and when it was used the doctor pushed the product and it broke.All pieces were successfully removed from the patient.No delay and no back up was available to complete the procedure.No patient injury was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10: the device reported, used in treatment, was not returned for evaluation.The relationship between the product and reported incident cannot be established.A review of manufacturing records for the reported lot number \ found no non-conformances or anomalies during manufacturing process related to the reported event.A complaint history review found no related failures; this failure mode will be trended to assess for any necessary corrective actions.A review of the product/print specifications found no discrepancies.Without the reported product a visual inspection and functional evaluation cannot be performed and customer´s complaint cannot be confirmed.Potential factors unrelated to the design or manufacture of the device that may lead to the failure reported include, but are not limited to: (1) excessive probing.(2) misalignment of implant during insertion.(3) shallow bone hole (4) over tensioning.No containment or corrective actions are recommended at this time.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
 
Additional Manufacturer Narrative
The multifix s ultra 6.5mm knotless anchor device, used in treatment, was returned for evaluation.A relationship between the device and reported incident was established.A review of manufacturing records for the reported lot number 2035022 found no non-conformances or anomalies during manufacturing process related to the reported event.A complaint history review found no related failures; this failure mode will be trended to assess for any necessary corrective actions.A review of the product/print specifications found no discrepancies.Visual evaluation shows the device has been returned deployed.The implant is not available.The device is intended for single use and can not be tested.The complaint was verified and the root cause could not be determined with certainty.Factors unrelated to the design and manufacture of the device which could have contributed to the complaint event are: (1) excessive force (2) over tensioning the sutures (3) bending or twisting the inserter handle during and after insertion.No containment or corrective actions are recommended at this time.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIX S-ULT 6.5MM KNOTLESS ANCHOR
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key9827516
Report Number3006524618-2020-00109
Device Sequence Number14174560
Product Code MBI
UDI-Device Identifier00885556614211
UDI-Public00885556614211
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K153669
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/11/2022
Device Model Number72290002
Device Catalogue Number72290002
Device Lot Number2035022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2020
Initial Date Received by Manufacturer 02/19/2020
Supplement Date Received by Manufacturer05/15/2020
06/18/2020
Initial Report FDA Received Date03/12/2020
Supplement Report FDA Received Date05/18/2020
06/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-