• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0308
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2020-00354.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure for a vessel sacrifice in the left vertebral artery using penumbra smart coils (smart coils) and non-penumbra microcatheter.During the procedure, the physician placed multiple smart coils in the target vessel using the microcatheter.The physician then attempted to advance another smart coil; however, the coil would not advance at all within its introducer sheath.Therefore, the smart coil was removed.The physician attempted to advance a new smart coil, however, the same issue occurred, and the coil would not advance within its introducer sheath; therefore, the smart coil was removed.The procedure was completed using new smart coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.The embolization coil was undamaged and intact with the pusher assembly.Conclusions: evaluation of the returned smart coil revealed that the pusher assembly mid-joint was retracted into the introducer sheath friction lock.If this occurs, resistance will likely be experienced while advancing the device.Forceful advancement of the device against this resistance may have contributed to the kinks observed near the pusher assembly mid-joint.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.The smart coil was then able to advance through its introducer sheath with resistance due to the kinks on the pusher assembly.The device was unable to be advanced through a demonstration microcatheter due to the pusher assembly kink.Evaluation of the returned smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced while advancing the device.During functional testing, the smart coil was unable to advance within its introducer sheath from its returned position within the introducer sheath; therefore, the introducer sheath was removed distally, and the smart coil was re-sheathed properly.The smart coil was then able to advance through its introducer sheath and a demonstration microcatheter without issue.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2020-00354.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9831178
Report Number3005168196-2020-00355
Device Sequence Number740379
Product Code HCG
UDI-Device Identifier00814548015798
UDI-Public00814548015798
Combination Product (Y/N)Y
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/13/2024
Device Model Number400SMTHXSFT0308
Device Catalogue Number400SMTHXSFT0308
Device Lot NumberF89077
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/19/2020
Supplement Date Received by Manufacturer03/23/2020
Initial Report FDA Received Date03/13/2020
Supplement Report FDA Received Date04/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient Weight90
Date Report Sent to FDA01/01/2005
-
-