| Model Number |
2429-0500 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
High Blood Pressure/ Hypertension (1908)
|
| Date of Event |
02/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a levophed infusion that the pump alarmed "occlusion".The nurse discovered that the silicone segment was occluded, sealing the sides of the tubing together which inhibited the correct flow of medication to the patient.The patient¿s blood pressure dropped and the levophed drip was increased to support the patient's blood pressure.There was no permanent harm.
|
| |
|
Event or Problem Description
|
|
It was reported that during a levophed infusion that the pump alarmed "occlusion".The nurse discovered that the silicone segment was occluded, sealing the sides of the tubing together which inhibited the correct flow of medication to the patient.The patient¿s blood pressure dropped and the levophed drip was increased to support the patient's blood pressure.There was no permanent harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report that the silicone segment was occluded, sealing the sides of the tubing together was confirmed.The set was visually inspected for obvious damage such as incomplete bonding engagements, cracks, kinks, holes, and tears in the tubing and its components.Visual inspection observed two fused sections in the silicone tubing.The first fused section was approximately 2.0 inches below the upper fitment.The second fused section was approximately 1.75 inches above the lower fitment.Although fluid did not initially fill the drip chamber nor flow through set, the fused sections of silicone tubing were massaged per suggestions from the chamber blocked tip sheet until they were no longer fused.Fluid was then observed to be able to flow through the set.The root cause of the this failure was not identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during a levophed infusion that the pump alarmed "occlusion".The nurse discovered that the silicone segment was occluded, sealing the sides of the tubing together which inhibited the correct flow of medication to the patient.The patient¿s blood pressure dropped and the levophed drip was increased to support the patient's blood pressure.There was no permanent harm.
|
| |
|