Section h3: additional information manufacturer's investigation conclusion: low flow events were confirmed based on the retrieved log file data.A direct correlation between (b)(6) and the reported ventricular tachycardia cannot be conclusively determined through this investigation.The retrieved left ventricular assist device (lvad) event log file contained relevant data spanning from (b)(6) 2020 at 19:28:32 to (b)(6) 2020 at 18:00:49, per the timestamp.From (b)(6) 2020 at 19:28:32 through (b)(6) 2020 at 14:46:23 average flow ranged from 1.7 to 4.4 liters per minute (lpm).Beginning on (b)(6) 2020 at 17:02:54, a decrease in flow was observed for the remainder of the log file, with average flow ranging from 0.0-1.5lpm.A specific cause for these low calculated flow values could not be conclusively determined through this evaluation.The device was returned assembled with the pump cable severed approximately 5¿ from the pump header.A severed portion of pump cable measuring approximately 13¿ was also returned.Approximately 3¿ of the sealed outflow graft was returned detached from the device.The apical cuff and modular cable were not returned.Examination of the pump's blood-contacting surfaces revealed depositions of clotted blood admixed with post-explant blood were observed in the inflow cannula and in the rotor vanes.These depositions appeared to have developed as a result of poor surface washing due to a low flow event or interruption in flow, consistent with the data captured in the lvad event log file.No adhered/developed depositions or thrombus formations were observed during the evaluation of the returned device.Visual inspection of the pump rotor and rotor well did not reveal any obvious surface scratches or defects.The device was cleaned, rebuilt, and tested under loaded conditions on a mock circulatory loop.The device functioned as intended and in accordance with manufacturing specifications.The heartmate 3 left ventricular assist system (lvas) instructions for use (ifu) lists cardiac arrhythmia as an adverse event, as well as a potential late postimplant complication that may be associated with the use of the heartmate 3 left ventricular assist system.The ifu explains that the low flow hazard alarm will be triggered when pump flow is less than 2.5lpm and notes that changes in patient conditions can result in low flow.The ifu, as well as the patient handbook, describe all system alarms and the recommended actions associated with them.No further information was provided.The manufacturer is closing the file on this event.
|