| Model Number |
PMX220 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 220 (pump max) and a penumbra system ace 68 reperfusion catheters (ace68).During the procedure, the pump max aspiration pressure would not increase beyond -20 inhg resulting in the decrease of suction force for the thrombus; therefore, the pump max was removed.It was reported that multiple attempts were made to troubleshoot the pump max but were unsuccessful.The procedure was completed using manual aspiration with a 60cc syringe.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Results: a pipe cleaner was inserted into the vacuum inlet, and blood was observed inside the tubing.Conclusions: evaluation of the returned pump max revealed blood within the tubing.If the aspiration tubing is connected directly to the vacuum inlet instead of a canister, supplied by penumbra, blood may be aspiration into the pump assembly.If this occurs, the pump max may not function properly.Penumbra pumps are 100% visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Search Alerts/Recalls
|