| Catalog Number |
306499 |
| Medical Device Problem Code |
Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Based on no sample, the investigation concluded, bd was not able to verify the indicated failure.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: this is the 1st complaint for lot # 9002664 for this type of defect or symptom.There was no documentation for this type of defect during the entire production run of this batch.Root cause description: undetermined.Rationale: capa not required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of syrs 10 ml fil saline shortlength plunger rods experienced syringe volumetric accuracy issues which were noted during use.The following information was provided by the initial reporter: material no: 306499 batch no: 9002664.Customer stated that the syringe is off by.4 -.6ml when aspirated with medication.He reported this happening before but said nothing came of it.He states this has been happening since november 2019.
|
| |
|
Search Alerts/Recalls
|