• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD RIGHT SIZE 8; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD RIGHT SIZE 8; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Pain (1994); Tissue Damage (2104)
Date of Event 01/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: associated products: item#:42540000035; all poly patella cemented 35 mm diameter; lot#:64189156.Item#:42530007902;natural tibia trabecular metal two-peg porous fixed bearing right size g; lot#:63905801.Item#:42522100912; articular surface medial congruent (mc) right 12 mm height use with tibia sizes g-h/cr femur sizes 8-11; lot#:64167122.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it is still implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920-2020-00175.
 
Event or Problem Description
It was reported that the patient underwent an initial total right knee arthroplasty and subsequently, he experienced right knee patellar clunk and received an intraarticular corticosteroid injection approximately one year later.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g4; g7; h1; h2; h3; h6.Reported event was unable to be confirmed due to limited information received from the customer.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Device history record (dhr) was reviewed and no discrepancies were found.Review of the available records identified the following: no inter-ops complication during the initial surgery.Crf review found patient was experiencing patella clunk and pain.The patient was administered with corticosteroid injection in the joint.A definitive root cause cannot be determined.Per package insert, persona the personalized knee system: pain is known adverse effect of this system.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD RIGHT SIZE 8
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key9834867
Report Number3007963827-2020-00080
Device Sequence Number2887057
Product Code MBH
UDI-Device Identifier00889024230385
UDI-Public(01)00889024230385
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42502606402
Device Lot Number63964229
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/21/2020
Supplement Date Received by Manufacturer06/12/2020
Initial Report FDA Received Date03/16/2020
Supplement Report FDA Received Date06/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Weight89
-
-