U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP,TSV,MCOL MG,4.7MM,16M; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP,TSV,MCOL MG,4.7MM,16M; DENTAL IMPLANT Back to Search Results
Model Number TSVMWB16
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Premarket identification pma/510(k) number k101880.
 
Event or Problem Description
It was reported that the implant tsvmwb16 failed due to bone loss in dental site 30.On the second stage appointment, it was noted bone loss around implant at crestal level, the implant was removed with a reversed torque.The patient will return for second stage for new implant, the site was grafted.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).The following sections have been updated: b4: date of this report.B5: describe event or problem.G4: date received by manufacturer.G7: type of report, follow-up number.H2: follow up type.H3: device evaluated by manufacturer.H6: event problem and evaluation codes.H10: additional narrative.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,MCOL MG,4.7MM,16M
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key9834898
Report Number0002023141-2020-00554
Device Sequence Number16528537
Product Code DZE
UDI-Device Identifier00889024019805
UDI-Public(01)00889024019805(17)200430(10)63047052(241)TSVMWB16
Combination Product (Y/N)N
Initial Reporter StatePR
Initial Reporter CountryUS
PMA/510(K) Number
K111889
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 05/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model NumberTSVMWB16
Device Catalogue NumberTSVMWB16
Device Lot Number63047052
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/19/2020
Supplement Date Received by Manufacturer05/15/2020
Initial Report FDA Received Date03/16/2020
Supplement Report FDA Received Date05/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight66
-
-