• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Vitreous Floaters (1866); Blurred Vision (2137); Visual Disturbances (2140)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information: date of event: between 12/16/2019 to 2/18/2020.Explanted, give date: not applicable as lens has not been explanted.(b)(4).Device evaluation: product testing could not be performed as the product was not returned.The reported complaint could not be verified.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zlb00 intraocular lens (iol) 20.0 diopter was implanted in the patient¿s right eye and a zkb00 20.0 was implanted in the left eye.The patient is experiencing halos, blurry vision and starbursts.Per the patient, he was told at the doctor¿s office before the surgeries that he¿d have the eyesight of a 20-year old.However, after surgery, patient is experiencing halos, blurred vision and starbursts.He¿s also having difficulty while driving at night.The sign posts look twice their actual size.The patient is not taking any eye drops.Thru follow-up the patient reported he had floaters prior to the implants.He still has floaters post implants but now the floaters are worse.He clarified, he is having the same symptoms in both eyes.The left eye is just worse.No further information was provided.This mdr will capture information for the right eye iol zlb00.The left eye is captured in a separate mdr.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9836343
Report Number9614546-2020-00124
Device Sequence Number16447206
Product Code MFK
UDI-Device Identifier05050474560673
UDI-Public(01)05050474560673(17)211118
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/18/2021
Device Model NumberZLB00
Device Catalogue NumberZLB00U0200
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/18/2020
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date03/16/2020
Supplement Report FDA Received Date11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Date Report Sent to FDA01/01/2005
-
-