• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION 1.5MM SYSTEM TWIST DRILL WITH J NOTCH 1.1 X 50MM, 7MM STOP; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION 1.5MM SYSTEM TWIST DRILL WITH J NOTCH 1.1 X 50MM, 7MM STOP; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).The product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source ¿ (b)(6).
 
Event or Problem Description
It was reported that a drill bit fractured during an open reduction and internal fixation of a zygomatic fracture.The drill bit fractured when it was applied to a fixation plate.X-rays were taken and the fractured piece of the drill could not be found in the patient.No adverse events were reported as a result of the malfunction.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is confirmed.The drill part number 01-7144 tw drill 1.1x50mm 7mmstop w/nt serial number (b)(6) was returned for evaluation.On visual inspection, the drill was noted to have a transverse fracture through the flutes near the neck.The cert was reviewed and there were no non-conformances were found.There are no indications of manufacturing defects.This is the only complaint in regards to the drill fracturing for part 01-7144 serial number (b)(6).For this part (01-7144) and the previous one year (from the notification date) regarding the drill fracturing, there is a complaint rate of (b)(4) which is no greater than the occurrence listed in the application fmea.The most likely underlying cause of the complaint is the drill experienced force in excess of what it was designed to encounter; fractures at this point of the drill are typical of excessive force fractures.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.The following fields were updated: b4 date of this report.B5 describe event or problem.D10 device availability.G4 date received by manufacturer.G7 type of report.H2 follow up type.H3 device evaluated by manufacturer.H4 device manufacturer date.H6 method code.H6 results code.H6 conclusions code.H10 additional narratives/data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5MM SYSTEM TWIST DRILL WITH J NOTCH 1.1 X 50MM, 7MM STOP
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9836490
Report Number0001032347-2020-00167
Device Sequence Number3037320
Product Code HBE
UDI-Device Identifier00841036013165
UDI-Public00841036013165
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062842
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number01-7144
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/06/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/27/2020
Supplement Date Received by Manufacturer07/29/2020
Initial Report FDA Received Date03/16/2020
Supplement Report FDA Received Date07/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-