| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).The product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source ¿ (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that a drill bit fractured during an open reduction and internal fixation of a zygomatic fracture.The drill bit fractured when it was applied to a fixation plate.X-rays were taken and the fractured piece of the drill could not be found in the patient.No adverse events were reported as a result of the malfunction.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The complaint is confirmed.The drill part number 01-7144 tw drill 1.1x50mm 7mmstop w/nt serial number (b)(6) was returned for evaluation.On visual inspection, the drill was noted to have a transverse fracture through the flutes near the neck.The cert was reviewed and there were no non-conformances were found.There are no indications of manufacturing defects.This is the only complaint in regards to the drill fracturing for part 01-7144 serial number (b)(6).For this part (01-7144) and the previous one year (from the notification date) regarding the drill fracturing, there is a complaint rate of (b)(4) which is no greater than the occurrence listed in the application fmea.The most likely underlying cause of the complaint is the drill experienced force in excess of what it was designed to encounter; fractures at this point of the drill are typical of excessive force fractures.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.The following fields were updated: b4 date of this report.B5 describe event or problem.D10 device availability.G4 date received by manufacturer.G7 type of report.H2 follow up type.H3 device evaluated by manufacturer.H4 device manufacturer date.H6 method code.H6 results code.H6 conclusions code.H10 additional narratives/data.
|
| |
|