| Model Number |
48132601 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
02/20/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is anticipated to be returned.
|
| |
|
Event or Problem Description
|
|
It was reported that a xia titanium 4.5 vitalium rod fractured post-operatively.Surgery was performed by glowing rod method, with the first extension scheduled for (b)(6) 2020.An examination of the patient revealed a broken rod before first extension could be performed.Revision was performed to replace the rod.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, functional and material analysis could not be performed as the device was not returned.However, x-rays were provided and confirmed that a rod on the right inferior side of the construct is fractured.The x-rays shows several levels of fixation at the top of the construct, and several levels of fixation at the bottom, and a section of non fixation in the middle.A review of device and complaint history could not be performed, as the lot number was not provided because the device was no returned.As the device was not returned for evaluation and additional information about the patient was not provided, a definite root cause cannot be determined.However likely root cause is the bending moment created by the construct and the existing curve of the spine caused the rod to fracture.
|
| |
|
Event or Problem Description
|
|
It was reported that a xia titanium 4.5 vitallium rod fractured post-operatively.Surgery was performed by growing rod method, with the first extension scheduled for (b)(6) 2020.An examination of the patient revealed a broken rod before first extension could be performed.Revision was performed to replace the rod.
|
| |
|
Search Alerts/Recalls
|