Model Number 242012
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Udi: (b)(4).
 
Event or Problem Description
It was reported by the sales rep via email that when 2 of the arthscp,4/17530jl/storz hub arrived had stains that could not be removed.No patient consequence and no surgical delay was reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the complaint device was received at the service center and evaluated.It was reported that the optics had stains that could not be cleaned.This complaint can be confirmed.It was identified that the tube was bent and the optics was damaged.The repair was declined and the device was discarded.User mishandling and lack of maintenance can be the most probable root causes of the failures.The potential cause for this issue is not related to manufacturing, therefore a manufacturing record evaluation is not required.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHSCP,4/175_30_JL/STORZ HUB
Common Device Name
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9836805
Report Number1221934-2020-00862
Device Sequence Number13840739
Product Code HRX
UDI-Device Identifier10886705026210
UDI-Public10886705026210
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K971996
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number242012
Device Catalogue Number242012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Initial Date Received by Manufacturer 03/02/2020
Supplement Date Received by Manufacturer11/02/2020
11/03/2020
Initial Report FDA Received Date03/16/2020
Supplement Report FDA Received Date11/02/2020
11/04/2020
Patient Sequence Number1
-
-