| Model Number |
PED-400-20 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pipeline flex pushwire was returned for evaluation.There was no pipeline flex braid returned with the pushwire.The pipeline flex pushwire was found outside of the phenom 27 catheter.The pipeline flex pushwire appeared to be detached at the hypotube proximal to the wire weld.No defects were found with the tip coil, distal marker, re-sheathing marker or with the proximal bumper.The ends of the detached pushwire were sent out for sem/eds analysis.Based on the analysis findings and sem/eds analyses, the pipeline flex was confirmed to have "pushwire detached at hypotube proximal to the wire weld".A review of the manufacturing process and lot history record did not uncover any deficiencies.In addition, the elemental analysis conducted through scanning electron micrographic (sem) / energy dispersive spectroscopy (eds) showed presence of soldering material (tin); thereby indicating that the soldering was conducted.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received report that the pipeline flex distal tip of the delivery system separated.The device separation occurred after deployment of the device and during recapture of the delivery wire.The separation occurred within the catheter.The device tip and the devilry system will be returned for analysis.There was no patient impact as a result of this event.The vessel anatomy was moderate.This event occurred in the left vertebral.A continuous flushed was used during the case.The devices were prepared and used per the instructions for use (ifu).Resistance was felt when the delivery wire was retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Information contained in mw5093626 was previously reported under manufacturer's report# 2029214-2020-00236.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|