| Model Number |
2100-12 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
We have received the complaint device for evaluation and we have confirmed the reported incident.When we attempted to inflate the balloon with saline, we observed a leakage at the distal bond between the balloon and the catheter shaft.The root cause of the issue was determined to be an operator error- not enough glue was applied on the joint during the assembly process.We have not received any other complaints of a similar nature for devices from this lot.Therefore, we believe that it was an isolated incident.Surgeon detected the malfunction during pre-use check.Device was not used for the procedure.He then used another pruitt aortic occlusion catheter for the procedure.
|
| |
|
Event or Problem Description
|
|
During pre-use check, surgeon observed a leakage at the tip of the balloon when he attempted to inflate the balloon with saline.Device was not used for the procedure.
|
| |
|